Reflectivity Measurement of Weak Fiber Bragg Grating (FBG)
来源期刊:Journal Of Wuhan University Of Technology Materials Science Edition2012年第6期
论文作者:郭会勇
文章页码:1177 - 1179
摘 要:One method based on the end-face reflection as a reference was used to measure weak grating reflectivity of 0.01%-1%. For measuring ultra-weak grating reflectivity, the grating group with an identical wavelength was used. By using anhydrous alcohol and deionized water as the reflecting medium, the single-pulse FBG reflectivity of a highly doped-Ge photosensitive fiber was calculated to be about 0.747% and 0.739%, respectively; and the single-pulse FBG reflectivity of ordinary SMF fiber was about 0.016% and 0.015%, respectively. The consistent results by the two different reflecting media showed that the end-face reflection method on the grating group was feasible.
郭会勇
National Engineering Laboratory of FOS Technology, Wuhan University of Technology
摘 要:One method based on the end-face reflection as a reference was used to measure weak grating reflectivity of 0.01%-1%. For measuring ultra-weak grating reflectivity, the grating group with an identical wavelength was used. By using anhydrous alcohol and deionized water as the reflecting medium, the single-pulse FBG reflectivity of a highly doped-Ge photosensitive fiber was calculated to be about 0.747% and 0.739%, respectively; and the single-pulse FBG reflectivity of ordinary SMF fiber was about 0.016% and 0.015%, respectively. The consistent results by the two different reflecting media showed that the end-face reflection method on the grating group was feasible.