An Unusually High Sensitivity of Pt-black pH Electrode
来源期刊:Journal Of Wuhan University Of Technology Materials Science Edition2010年第2期
论文作者:孙咏红
文章页码:197 - 199
摘 要:The potentials of Pt-black electrode using a copper conducting wire instead of the salt bridge in acid and alkaline solutions without the use of H2 evolution reactions were measured. There were three nonlinear portions in the calibration curve. Unusually, the potential slopes at pH 3-5 and 8-10 indicated 200 mV and 70 mV per pH, respectively. Such high sensitivity for pH slope, more than 4 times of usual 59 mV per pH, may be credited to the special properties of the Pt-black surface. SEM (scanning electronical microscopy) was applied to characterize the surface of the Pt-black electrode. Its working mechanism is well explained in the theory of capacitance potentials rather than Nernst’s redox potentials.
孙咏红
School of Environmental and Chemical Engineering, Dalian Jiaotong University
摘 要:The potentials of Pt-black electrode using a copper conducting wire instead of the salt bridge in acid and alkaline solutions without the use of H2 evolution reactions were measured. There were three nonlinear portions in the calibration curve. Unusually, the potential slopes at pH 3-5 and 8-10 indicated 200 mV and 70 mV per pH, respectively. Such high sensitivity for pH slope, more than 4 times of usual 59 mV per pH, may be credited to the special properties of the Pt-black surface. SEM (scanning electronical microscopy) was applied to characterize the surface of the Pt-black electrode. Its working mechanism is well explained in the theory of capacitance potentials rather than Nernst’s redox potentials.
关键词: